Design of a self-aligned, wide temperature range (300 mK-300 K) atomic force microscope/magnetic force microscope with 10 nm magnetic force microscope resolution


Karci O., Dede M., ORAL A.

REVIEW OF SCIENTIFIC INSTRUMENTS, cilt.85, sa.10, 2014 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 85 Sayı: 10
  • Basım Tarihi: 2014
  • Doi Numarası: 10.1063/1.4897147
  • Dergi Adı: REVIEW OF SCIENTIFIC INSTRUMENTS
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Hacettepe Üniversitesi Adresli: Hayır

Özet

We describe the design of a wide temperature range (300 mK-300 K) atomic force microscope/magnetic force microscope with a self-aligned fibre-cantilever mechanism. An alignment chip with alignment groves and a special mechanical design are used to eliminate tedious and time consuming fibre-cantilever alignment procedure for the entire temperature range. A low noise, Michelson fibre interferometer was integrated into the system for measuring deflection of the cantilever. The spectral noise density of the system was measured to be similar to 12 fm/root Hz at 4.2 K at 3 mW incident optical power. Abrikosov vortices in BSCCO(2212) single crystal sample and a high density hard disk sample were imaged at 10 nm resolution to demonstrate the performance of the system. (C) 2014 AIP Publishing LLC.