Comparison of shear bond strengths of orthodontic brackets bonded with flowable composites


Creative Commons License

TURGUT M. D., ATTAR N., Korkmaz Y., Gokcelik A.

DENTAL MATERIALS JOURNAL, vol.30, no.1, pp.66-71, 2011 (SCI-Expanded) identifier identifier identifier

  • Publication Type: Article / Article
  • Volume: 30 Issue: 1
  • Publication Date: 2011
  • Doi Number: 10.4012/dmj.2010-102
  • Journal Name: DENTAL MATERIALS JOURNAL
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Page Numbers: pp.66-71
  • Hacettepe University Affiliated: Yes

Abstract

This study evaluated the shear bond strengths of orthodontic brackets bonded to human premolars using five different combinations of flowable composites and one-step self-etching adhesives (n=12): (1) Adper Easy Bond+Filtek Supreme XT Flow; (2) Futurabond NR+Grandio Flow: (3) Clearfil S3 Bond+Clearfil Majesty Flow; (4) AdheSE One+Tetric EvoFlow; and (5) Transbond Plus Self Etching Primer+Transbond XT Light Cure Adhesive. After shear bond strength testing, adhesive remnant index (ARI) scores were given according to the amount of adhesive and resin remaining on the brackets. On shear bond strength, there were no statistically significant differences between Groups 2 and 4 and between Groups 3 and 5 (p > 0.05). On ART scores, the predominant ARI scores in Groups 1, 2, 3, and 5 were 4, 2, 5, and 4 respectively: in Group 4, they were 0 and 4. Results showed that some combinations of flowable composites and self-etching adhesives might not he suitable for orthodontic use due to their low shear bond strengths and high ARI scores-with the latter signaling the risk of damaging the enamel surface during debonding.