Electrochemical Detection of Cortisol by Silver Nanoparticle-Modified Molecularly Imprinted Polymer-Coated Pencil Graphite Electrodes


Shama N. A., Aşır S., GÖKTÜRK BAŞAL I., Yılmaz F., TÜRKMEN D., DENİZLİ A.

ACS Omega, cilt.8, sa.32, ss.29202-29212, 2023 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 8 Sayı: 32
  • Basım Tarihi: 2023
  • Doi Numarası: 10.1021/acsomega.3c02472
  • Dergi Adı: ACS Omega
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Directory of Open Access Journals
  • Sayfa Sayıları: ss.29202-29212
  • Hacettepe Üniversitesi Adresli: Evet

Özet

The sensitive cortisol detection by an electrochemical sensor based on silver nanoparticle-doped molecularly imprinted polymer was successfully improved. This study describes the method development for cortisol detection in both aqueous solution and biological samples using molecularly imprinted poly(hydroxyethyl methacrylate-N-methacryloyl-(l)-histidine methyl ester)-coated pencil graphite electrodes modified with silver nanoparticles (AgNPs) by differential pulse voltammetry (DPV). The cortisol-imprinted pencil graphite electrode (PGE) has a large surface area because of doped AgNPs with enhanced electroactivity. The prepared molecularly imprinted polymer was characterized by scanning electron microscopy. The DPV response of the synthesized electrode with outstanding electrical conductivity was clarified. Cortisol-imprinted polymer-coated PGEs (MIP), cortisol-imprinted polymer-coated PGEs with AgNPs (MIP@AgNPs), and nonimprinted polymer-coated PGEs with AgNPs (NIP@AgNPs) were evaluated for sensitive and selective detection of cortisol in aqueous solution. Five different cortisol concentrations (0.395, 0.791, 1.32, 2.64, and 3.96 nM) were applied to the MIP@AgNPs, and signal responses were detected by the DPV with a regression coefficient (R2) value of 0.9951. The modified electrode showed good electrocatalytic activity toward cortisol for the linear concentration range from 0.395 to 3.96 nM, and a low limit of detection was recorded as 0.214 nM. The results indicate that the MIP@AgNPs sensor has great potential for sensitive and selective cortisol determination in biological samples.