The association between urinary BPA levels and medical equipment among pediatric intensive care patients
ENVIRONMENTAL TOXICOLOGY AND PHARMACOLOGY, cilt.83, 2021 (SCI-Expanded, Scopus)
- Yayın Türü: Makale / Tam Makale
- Cilt numarası: 83
- Basım Tarihi: 2021
- Doi Numarası: 10.1016/j.etap.2021.103585
- Dergi Adı: ENVIRONMENTAL TOXICOLOGY AND PHARMACOLOGY
- Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Academic Search Premier, BIOSIS, CAB Abstracts, Chemical Abstracts Core, EMBASE, Environment Index, Greenfile, MEDLINE, Pollution Abstracts, Veterinary Science Database
- Hacettepe Üniversitesi Adresli: Evet
Özet
We aim to evaluate urinary total BPA (tBPA) levels and association with medical devices used on patients in pediatric intensive care units. This cross-sectional descriptive study included 117 critically ill children. Urinary tBPA levels were determined using high-performance liquid chromatography. General estimating equations with repeated measures analyzed the effect of interventions and devices on urinary BPA levels. A total of 292 urine samples taken from 117 child intensive care patients were studied. When age, sex, and body mass index-for age z scores were controlled, cases having endotracheal intubation showed higher urinary tBPA levels (p = 0.003) and hemodialyzed patients had considerably higher urinary tBPA levels (p = 0.004). When confounding factors were controlled, cases using both multiple iv treatment and more than four medical devices showed higher urinary tBPA levels than their counterparts (p = 0.007 and p = 0.028, respectively). The use of certain medical devices and interventions could increase BPA exposure in pediatric intensive care patients.